Delete values
automatically (AClr)
98 •
5. To change the setting or call up another submenu,
keep pressing the arrow key send or print until the
required setting (here: level "2") is displayed.
6. Confirm the setting with the Enter key (input).
You will leave the menu automatically.
7. To make further settings, call up the menu again
and repeat the process.
NOTE
• Press the menu key briefly to go back one
menu level.
• Press and hold the menu key to exit the
menu at any time.
• If no key is pressed for approx. 24 seconds,
the menu is automatically quit.
Out-of-date measured results and patient data lead to
incorrect calculation of BMI/BFR. You can set the
device so that the following measured results and
patient data are cleared automatically after 5 minutes:
• Gender
• Physical activity level (PAL)
• Age
• Height
• BMI
• BFR
NOTE
• If you want to enter PAL, age and gender
(input function) for the next measurement, the
values of the last measurement are proposed.
(see "Enter patient data (input)" on page 91).
• With certain models, this function is already
activated at the factory. If required, you can
disable this function.
1. Select the item "AClr" from the menu.
2. Confirm the selection.
The current setting is displayed.
3. Select the setting you require
– On
– Off