5.6 Calibrate head slide

Automated calibration

CAUTION!
Risk of injuries and damage to the device
The device must be tilted. The large overall
height of the device can result in injuries and
damage to the device.
− Ensure that there are no other persons in the
immediate vicinity.
− Ensure that there are no objects in the
immediate vicinity.
1. Tilt the device backwards.
ATTENTION!
Damage to the device due to excessive voltage
Standard power supplier units can supply a
higher voltage than is given on them. The scales
can overheat, catch fire,
melt or be short-circuited
− Only use genuine seca plug-in power supply
units with 9V or a controlled 12 V output
voltage.
2. Insert the connector plug of the power supply unit
in the connecting socket of the scales.
3. Carefully upright the device.
4. Plug the power supply unit into a mains socket.
Before you can measure height using the device for the
first time, you have to calibrate it. You do so by
"teaching" the device a reference measurement.
You can perform automated calibration using the
calibration rod supplied.
1. Press the Start key
"----" appears in the display.
of the head slide.
Before you get started ... • 85