Inscriptions and Symbols on the Unit
10 A
Input for measuring current in the ampere measuring area up to
10 A - AC/DC. In the 10 A range the measuring time must be limited to
a maximum of 15 seconds and the next measuring should happen no
sooner than after 15 minutes. This measuring area is not protected
with a fuse. If the current is higher than 10 A, the unit can be
destroyed.
A
Input for measuring current of up to 2 A - AC/DC. The Input is
protected with a fuse (F2A/250V Ø5x20 mm).
~
AC
DC
Dangerously high voltage between the inputs. Use extreme caution
when taking a measurement. Do not touch the inputs and the metal
tips of the probes.
Attention! Please read the manual carefully!
Protection Class II
Connecting the Measuring Cables
The measuring cables provided are capable of taking measurements up to a
maximum of 1000 V.
The measurement of high voltages should always be performed with great
care and only when a person trained in first aid is present.
Important!
The maximum allowable input voltage is 300 V (CAT
I)
1000 V (CAT
Using the Stand
The unit is fitted with a stand integrated into the rubber protector for propping it
up on a table or similar. To use the stand, grip the lower end of the flap and
pull outwards.
Before Use
and may not be exceeded.
III),
20
II
600 V (CAT
) resp.