Recommended separation distances
between portable and mobile RF communications equipment and the eQuinox
The eQuinox is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or
the user of the eQuinox can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the eQuinox as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum output
power of transmitter
W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be esti-
mated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from struc-
tures, objects and people.
24 — 4805-C1 C (English)
Separation distance according to frequency of transmitter
150 khz to 80 Mhz
d
1
2 .
P
0.12
0.38
1.2
3.8
12
m
80 Mhz to 800 Mhz
d
1
2 .
P
0.12
0.38
1.2
3.8
12
800 Mhz to 2,5 Ghz
d
1
2 .
P
0.23
0.73
2.3
7.3
23