2.4

Use in Ex-zones

Note!
For intended installation in an Ex-classified area, please observe the fol-
lowing guidelines:
Ex- classification:
Marking and number of the test certificate:
SEV 14 ATEX 0129 X
Introduction
1.
2.
3.
InPro 7100-VP Series
52 002 092
II 1/2 G Ex ia IIB T3/T4/T5/T6 Ga/Gb (plastic sensors)
II 1/2 G Ex ia IIC T3/T4/T5/T6 Ga/Gb (metallic sensors/metallic
sensors with plastic surface)
The conductivity sensors made of plastic are according to Directive
94/9/EC (ATEX 95) Annex I equipment group II, category 1/2G after
RL 99/92/EC (ATEX 137) in the zones 0/1 or 0/2 or 1/2 and the gas
groups IIA and IIB, explosion hazards due to flammable substances
in the field of temperature classes T1 to T6, may be used.
The conductivity sensors made of metal or metal coated with plastic
to Directive 94/9/EC (ATEX 95) Annex I to equipment group II, cate-
gory 1/2G after RL 99/92/EC (ATEX 137) in zones 0/1 or 0/2 or 1/2
and gas groups IIA, IIB and IIC, explosion hazards due to flammable
substances in the field of temperature classes T1 to T6, may be used.
For use/installation, the requirements according to EN 60079-14
must be observed.