Herunterladen Inhalt Inhalt Diese Seite drucken

Mettler Toledo InPro 7108-VP/ CPVC Bedienungsanleitung Seite 6

4-pol-leitfähigkeits-messzellen inpro 7100-vp series
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

6
4-electrode sensors are so-called diffuse-field sensors. The individually
determined cell constant was established under laboratory conditions
without external influences.
Depending on the installation configuration, external influences (e.g. from
the pipe) can result in a shift in cell constant. If the application does not
guarantee a minimum clearance of 30 mm (1.18") around the sensor, it is
recommended to carry out recalibration of the sensor in installed position.
Packaging
The packaging consists of cardboard and plastic material.
Keep the packaging for later use during storage or for transportation of
the sensor. Should you wish to dispose of the packaging material, please
observe your local regulations as well as those data and instructions given
in Chapter 5.3 of this manual.
Unpacking and inspection
Please check the sensors immediately during unpacking in order to determi-
ne possible damage or missing items. Any irregularities should immediately
be reportet to your carrier and to your supplier.
Each InPro 7100-VP Series sensor is equipped with an integral tempera ture
probe Pt 1000, IEC 751 Class A.
The nominal cell constant is 0.25 cm
sensor label. For each sensor, the cell constant and temperature constant
have been measured/established individually and the values have been
documented in the accompanying Quality Certificate. All calibrations can
be traced back to NIST and /or ASTM Standards.
Both the measuring range and the system measurement accuracy strongly
depend upon which type of transmitter is employed.
InPro 7100-VP Series
52 002 092
. The exact value is printed on the
–1
© 02 / 2017 Mettler-Toledo GmbH, CH - 8606 Greifensee
InPro 7100-VP Series
Printed in USA

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis