Table 8 – Signal input/output parts PORT (according to IEC 60601-1-2)
Phenomenon
Basic EMC standard
ELECTROSTATIC DISCHARGE
IEC 61000-4-2
Electrical fast transients/bursts
IEC 61000-4-4
Surges
IEC 61000-4-5
Line-to-ground
Conducted disturbances induced
IEC 61000-4-6
by RF fields
Electromagnetic
Compatibility
(EMC) Information
Professional healthcare
facility environment
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
±1 kV
100 kHz repetition frequency
±2 kV
3 V
0.15 MHz – 80 MHz
6 V in ISM bands between
0.15 MHz and 80 MHz
80 % AM at 1 kHz
Indicaciones sobre
compatibilidad
electromagnética (CEM)
IMMUNITY TEST LEVELS
HOME HEALTHCARE
ENVIRONMENT
Not applicable since released only
for professional healthcare facility
environment