Herunterladen Inhalt Inhalt Diese Seite drucken

Storz TL 300 Gebrauchsanweisung Seite 53

Kaltlicht-fontane power led 300
Inhaltsverzeichnis

Werbung

45
Hinweise zur
elektromagnetischen
Verträglichkeit (EMV)
Basic EMC standard or
Phenomenon
ELECTROSTATIC DISCHARGE
IEC 61000-4-2
Radiated RF EM fields
IEC 61000-4-3
Proximity fields from RF
IEC 61000-4-3
wireless communications
equipment
RATED power frequency
IEC 61000-4-8
magnetic fields
Electromagnetic
Compatibility
(EMC) Information
Table 4 – ENCLOSURE PORT (according to IEC 60601-1-2)
Professional healthcare
test method
facility environment
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
3 V/m
80 MHz – 2.7 GHz
80 % AM at 1 kHz
See table 9
30 A/m
50 Hz or 60 Hz
Indicaciones sobre
compatibilidad
electromagnética (CEM)
IMMUNITY TEST LEVELS
HOME HEALTHCARE
ENVIRONMENT
Not applicable since released only
for professional healthcare facility
environment

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis