Note
The temperatures T
chapter 5.3).
Example
(measured) density at 26 °C (T
(measured) density at 15 °C (T
α = ( 0.7937 – 1) / (26 – 15)
0.7844
α = (1.011856 – 1) / 11 = 0.011856 / 11 = 0.001078
1000 = 1.078; enter this value into the instrument.
α
x
10 temperature-compensation coefficients can be saved.
Input of temperature-compensation coefficients via Comp. Density.
For typical temperature-compensation coefficients, see Chapter 13.2.
5.3

Temperature Unit (Temp. Unit)

Data in °C or °F (selectable).
5.4

Measure Mode

Configuration of sample identification, stability mode and method of data storage.
Sample Name
Stability
Auto
Manu
Mode
Labo
Field
Custom User-defined setting.
METTLER TOLEDO Densito 30PX
and T
have to be entered in the selected unit (°C or °F, see
1
2
) 0.7844 g/cm
1
) 0.7937 g/cm
2
Sample identification.
A letter (a...z or space) can be set for the identification of
samples.
Stability control.
The result is accepted automatically when the display is stable.
Result accepted by pressing the ok/meas. key.
Method of data storage.
By pressing the ok/meas. key the result is saved and trans-
ferred (printer, PC).
By pressing the ok/meas. key the result is saved.
Memory in
Memory out
3
3
Save results.
Auto
Save result automatically.
Manu
Result saved by pressing the
ok/meas. key.
Transfer result to PC or printer.
Auto
Transfer result automatically.
Manu
Transfer result by pressing the
key.