Errors and warning messages
Fluctuations in measurement values
Was the same point on the sample
used for all measurements?
Yes. It may help to test the
calibration with an additional
standard if one is available.
Calibration correct:
Is the test surface completely even
and does the measuring device
have good contact with the
sample?
Yes: Device defective
Do not attempt to make any repairs yourself! If a malfunction occurs on
your measuring device, our Customer Service department will be happy
to help you as quickly as possible.
No. Check how high the
deviations are on the sample itself.
Calibration not correct:
Recalibrate and clean the standard
if necessary (Section 14)
No: In this case, major deviations
Please contact our Customer Ser-
vice department.