Calibration and Autodiagnosis
14
Autodiagnosis general lasts for
about 2 seconds. "Please clean
standard" or "Please test standard"
may be displayed. For more
information on cleaning, see page
47.
A message will appear in the
display informing you that
autodiagnosis has been completed
successfully.
In some cases, the system may
suggest you perform calibration
again. The reason for this may
often lie in ambient conditions. It is
also possible, however, that the
standard still has small amounts of
residue left over from cleaning. This
problem can generally be alleviated
by cleaning with a dry optical
cleaning cloth.
Calibrate
You should recalibrate the device if
ambient conditions have changed.
This includes especially changing
location if major changes in
temperature and relative humidity
may be expected as a result (for
example inside/outside).
When moving from cold areas to
warm areas, there is also a danger
of condensation. For this reason,
after there has been a change in
ambient conditions, you should
wait for an appropriate amount of