Filter change

In normal operating conditions, the compressor filter must be
replaced after every 200 operating hours, but at least once a
year.
Also please check the filter at regular intervals (after every
10 applications). If it is soiled (grey or brown colour) or blocked,
replace it. The filter must also be replaced with a new one if it
has become damp.
The filter cannot be cleaned and then reused!
NOTICE
Only use PARI air filters that are intended for use with your
compressor, otherwise your compressor might be dam-
aged.
To replace the filter, proceed as follows:
• Unscrew the filter holder from
your compressor using a suit-
able coin.
• Pull the old filter out of the filter
holder and fit the new filter in
its place.
• Reinsert the filter holder together with the new filter in your
compressor and screw it tight with the coin.
Servicing
The compressor must only be serviced by the PARI GmbH
Technical Service department or a service centre expressly au-
thorised by PARI. If the compressor is opened or manipulated
by anyone else, all claims under the warranty shall be void. In
these cases, PARI GmbH will also accept no liability.