Maintain the following times when disinfecting:
Wattage
850 – 950
600 – 850
Information about the amount of water required and about the
disinfection process is included in the instructions for use of the
devices in question (disinfector and microwave).
CAUTION
The product components must not be exposed to microwave
radiation unless they are in a microwave disinfector. If they
are heated in a microwave oven without the protection of the
disinfector, the individual parts will be damaged.
Do not use the "grill" function because this too will damage
the individual parts.
At the end of the disinfection process, the parts may still be
very hot. Therefore, it is essential to wait for the cooling time
to pass before taking the parts out, otherwise there is a risk
of burns.
Inadequate or ineffective disinfection encourages the growth
of bacteria and thus increases the risk of infection. Disin-
fection is not completed properly unless both the specified
minimum disinfection time and the required power output of
the microwave are observed. The disinfector must also be
perfectly clean. Check regularly that the microwave and the
disinfector are functioning perfectly.
Use only water for disinfection, never use chemicals.
– 70 –
Disinfection time
4 min.
6 min.
Cooling time
2 min.
2 min.