5. Configuring instrument
If no button is pressed for 3s while in
the configuration mode, the instrument
changes to the next setting option!
The TPM limits can be between 4 and
40%. However, the upper limit value
(A A - - H H i i ) has to be at least 1% higher
1 Switch on instrument and keep
- The instrument switches to the
configuration mode. A A - - H H i i and the upper
limit set for TPM lights up
Setting upper alarm limit (A A - - H H i i )
2 Set upper limit value using
3 Do not press any button for 3s.
- The display changes: A A - - L L o o and the
bottom set limit value for TPM appears
14
than the bottom limit (A A - - L L o o ).
and
pressed during the
display tests
.
.
(keep button pressed to fast-forward).
or
.
5. Configuring instrument
Setting the bottom alarm limit (A-Lo)
4 Set bottom limit using
(keep button pressed to fast-forward).
5 Do not press any button for 3s.
- The display changes: The settings for the
Alarm LED (on or 0FF) and
(
/
).
Switching Alarm LED on/off
6 Switch LED on or off (on , oFF) via
or
.
7 Do not press any button for 3s.
- The display changes: The set tempe-
rature unit (°C or °F) lights up (
Setting temperature unit
8 Set required temperature unit °C or °F via
or
.
9 Do not press any button for 3s.
- Instrument switches to measurement
mode
, the set values were accepted.
or
appear
/
).
15