There are 4 screw terminals and a grounding terminal in the rear
cover for connecting the electrical leads.
The screw terminals accept gauges up to 1.5 mm
accessible after removing the cover.
Screw terminals
Ground terminal
Cable glands
Fig. 5. KINAX SR 709 with screw terminals and cable glands.
Mounting:
Weight:
Regulations
Electromagnetic
compatibility:
Intrinsically safe:
Test voltage:
Housing protection:
Impulse voltage withstand:
Permissible common-
mode voltage:
Camille Bauer
2
and are
Rear
(cover)
Cover
Directly
or
Fixing with NAMUR mounting part
(see Fig. 9 and 10)
Approx. 1.1 kg
The standards DIN EN 50 081-2 and
DIN EN 50 082-2 are observed
Acc. to EN 50 020: 1994
2.2 kVeff, 50 Hz, 1 min. between...
... power supply and housing
... power supply and measuring
output
(with DC/AC power supply, with
electric isolation, with AC/DC power
pack)
500 Veff, 50 Hz, 1 min.
all electric connections to housing
(with DC power supply, without
electric isolation)
IP 54 acc. to EN 60 529
1 kV, 1.2/50 µs, 0.5 Ws
IEC 255-4, Cl. II
100 V, 50 Hz
Transmitter for position feedback
Environmental conditions
Climatic rating:
Altitude:
Transportation and
storage temperature:
KINAX SR 709
Standard version
Temperature – 25 to + 70 °C
Annual mean relative humidity ≤ 90%
or
Version with improved climatic
rating
Temperature – 40 to + 70 °C
Annual mean relative humidity ≤ 95%
Ex version
Temperature – 40 to + 60 °C at T6
resp. – 40 to + 75 °C at T5
Annual mean relative humidity ≤ 95%
2000 m max.
– 40 to 80 °C