13. Maintenance and Testing
A. Cleaning and Disinfection
See the DV5HH Instruction Guide (A-DV5HH) for information
on cleaning the humidifi er chamber and heater base.
The DV5HH Standard Heated Humidifi er is single patient use
only and therefore does not require disinfection.
B. Maintenance
The DV5HH Standard Heated Humidifi er System does not
require regular maintenance.
C. Testing
Since the DV5HH device is powered and controlled by the PAP,
it is important to determine which unit is at fault when there
is an issue with humidity/heat. Use the following procedures
to determine whether the DV5HH device or the PAP device is
faulty.
1. Connect the DV5HH device to a known good PAP device.
2. After turning the system on,
• if the LCD display does NOT show "Heat:XXX" on the
bottom line, proceed to the troubleshooting section.
• if the DV5HH device does NOT produce heat, proceed
to the troubleshooting section.
3. Connect the PAP to a known good DV5HH device, if the
LCD display does NOT show "Heat:XXX" on the bottom
line, proceed to the troubleshooting section.
If you detect excessive air or water leaks in the DV5HH sys-
tem, proceed to the troubleshooting section.
24
LT-2026
DV5HH