Alternating current (A/AC)
Range
200.0 µA
20.00 mA
200.0 mA
10.00 A
600 V overload protection
Fuses: µA/mA = 2x 160 mA resettable PTC fuses
10 A = High-performance FF12AH1000V ceramic fuse
≤ 6 A = continuous measurement, >6A = max. 10 secs at intervals of 15 minutes
The multimeter may display 3 counts when a measuring input is open
TrueRMS peak (Crest Factor (CF)) ≤3 CF over the entire range
TrueRMS peak for non-sinusoidal signals plus tolerance
CF >1.0 - 2.0
+ 3%
CF >2.0 - 2.5
+ 5%
CF >2.5 - 3.0
+ 7%
Resistance
Range
200.0 Ω*
2.000 kΩ*
20.00 kΩ
200.0 kΩ
2.000 MΩ
20.00 MΩ
600 V overload protection
Measuring voltage: Approx. 1.0 V, measuring current approx. 0.7 mA
*Accuracy for measuring range ≤ 200 Ω was calculated after deducting lead resistance from the REL function
Battery test
Range
1.5 V
9 V
600 V overload protection, resettable PTC fuse
62
Resolution
0.1 µA
0.01 mA
0.1 mA
0.01 A
Resolution
0.1 Ω
0.001 kΩ
0.01 kΩ
0.1 kΩ
0.001 MΩ
0.01 MΩ
Load resistance
Approx. 15 Ω
Approx. 1 kΩ
Accuracy
±(1.3% + 7)
±(2.6% + 7)
Accuracy
±(1.3% + 3)
±(1.0% + 8)
±(1.6% + 4)
±(3.0% + 6)
Resolution
0.001 V
0.01 V
Accuracy
±(5% + 3)