Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
English
IEC 60601 test
Compliance
level
level
3 Vrms
150 kHz to 80
MHz
Vrms
3 V/m
80 MHz to 2,5
V/m
GHz
Electromagnetic environment –
guidance
Portable and mobile RF
communications equipment should be
used no closer to any part of the
+
ClearReader
, including cables, than
the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
d = 1,2
P
P
d = 1,2
80 MHz to 800 MHz
P
d = 2,3
800 MHz to 2,5 GHz
whereP is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation distance in
metres (m).
+