• Place the product components
in the disinfector with enough
space between them to allow
the steam to reach all surfaces
reliably.
Maintain the following times when disinfecting:
Wattage
850 – 950
600 – 850
Information about the amount of water required and about the
disinfection process is included in the instructions for use of the
devices in question (disinfector and microwave).
– 91 –
Disinfection time
4 min.
6 min.
Cooling time
2 min.
2 min.