Replacing the filter

NOTICE
Danger of device fault due to incorrect air filter
If air filters that are not designed for the compressor are
used, the compressor may be damaged.
• Only use air filters provided by PARI for use with your
compressor.
In normal operating conditions, the compressor filter must be
replaced after every 200 operating hours, but at least once a
year.
Also please check the filter at regular intervals (after every
10 applications). If it is soiled (grey or brown colour) or blocked,
replace it. The filter must also be replaced with a new one if it
has become damp.
The filter cannot be cleaned and then reused!
To replace the filter, proceed as follows: