Guidance and Manufacturer's Declaration – Electromagnetic Immunity
Loop
™
The customer or user should ensure that it is used in such an environment.
Immunity
IEC 60601 Test
Test
Conducted RF
150 kHz to 80
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
80 MHz to 2.7 GHz
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which Loop™ is used exceeds
the applicable RF compliance level above, Loop
performance is observed, additional measures may be necessary, such as re-orienting or relocating Loop
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
is intended for use in the electromagnetic environment specified below.
Compliance Level
Level
3 Vrms
150 kHz to 80 MHz
MHz
10 V/m
10 V/m
80 MHz to 2.7 GHz
Electromagnetic Environment Guidance
Portable and mobile RF communications
equipment should be used no closer to any
part of Loop
recommended separation distance calculated
from the equation applicable to the frequency
3 Vrms
of the transmitter
Recommended separation distance
d = [ 3.5/V ] √P
d = [ 3.5/V ] √P 80 MHz to 800 MHz
d = [ 3.5/V ] √P 800 MHz to 2,7 GHz
Where P is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in meters
(m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance level in each
frequency range. Interference may
occur in the vicinity of equipment
marked with the following symbol.
should be observed to verify normal operation. If abnormal
™
27
, including cables, than the
™
.
™