4.2.2 Tack Mode

Tack mode is recommended to deliver a short burst of light (1,000 mW/cm
tacking adhesives. Tack mode does not use the closed-loop feature.
Press and hold the Start/Stop button for about 2 seconds to initiate Tack mode. This can be done from any Idle
screen, regardless of the irradiance and time settings shown on the screen.
Tack mode recommended technique:
1. Before starting a tack cycle position the lens of the handpiece within 3 to 4mm of the center of the targeted surface.
2. Activate the tack cycle by pressing and holding the Start/Stop button for about 2 seconds. The device will beep,
and the tack cycle will begin immediately.
3. Hold the light over the target surface for the 3 second Tack cycle.
4. After the tack cycle is complete, the screen will display the energy delivered, 3J (joules).
5. The screen will automatically return to Direct Restorative mode and the previously used settings will be shown on
the Idle screen.

4.3 SETTINGS

Loop
allows the user to custom configure several options in Settings including:
• Sound ON/OFF
• Information screen: Manufacturing Lot Number
How to enter or exit Settings:
Menu:
Press to cycle through the main menu of settings.
Select:
Press to select setting options.
Settings Steps:
To enter Settings and configure the curing light, press and release both the Menu button and Select button
simultaneously. This will allow Loop
Press the Menu button to scroll through the menu options and press the Select button to change the setting options.
After making a selection, press the Menu button and the selection is automatically saved.
To exit Settings, press and release both the Menu button and Select button simultaneously.
NOTE: Your selections will save automatically when you exit Settings.
Protective Barrier Sleeve
ON
OFF
Sound Level
ON
OFF
+
(press and release both buttons simultaneously)
to enter Settings, displayed with all blue circle icons.
Protective Barrier Sleeve ON/OFF (preset to ON)
Press the Select button to configure Loop
Barrier Sleeve. If this setting is changed, it is recommended to do a calibration (refer
to Section 4.4 "Calibration")
Sound ON/OFF (preset to ON)
Set Sound to ON or OFF (mute).
for 3 seconds) to the dental material for
2
Protective barrier sleeve ON/OFF
14