Disposal
The curing light must not be disposed of as normal household waste. Dispose unserviceable
batteries and curing lights according to the corresponding legal requirements in your country.
Batteries must not be incinerated.

4.11 USE WITH A RADIOMETER

The Loop
LED Curing Light system functions as an internal radiometer that ensures an accurate, calibrated energy
output. However, if you wish to test the handpiece on an external radiometer, operate the handpiece in Direct
Restorative mode, with closed-loop turned off. To observe the closed-loop function on a radiometer, operate in
Direct Restorative mode by first touching the lens to the radiometer surface before raising it to a distance to observe
the closed-loop functionality.

5. Maintenance and Cleaning

5.1 CLEANING DURING USE

The Loop
handpiece, charging base and light shield are not autoclavable and no portion can be sterilized. Only use
approved disinfecting solutions. Refer to Section 5.2 "Cleaning After Use".
To keep the Loop
handpiece clean and functioning properly, a new barrier sleeve should be used for each patient.
Only use Loop
barrier sleeves that are specifically designed for use with the Loop
The supplied Loop
lens cleaning cloth should be used exclusively for drying the lens after cleaning.

5.2 CLEANING AFTER USE

Only use approved disinfecting solutions. If using a spray, do not spray the disinfecting solution directly on the
device. Instead, spray or moisten a gauze or soft cloth with disinfecting solution and use it to wipe down and clean
the unit. This ensures that significant amounts of disinfectant solution do not wick into the seams of the unit. When
finished, dry any residual disinfectant solution remaining on the surface of the handpiece with a soft cloth. Do not use
the supplied Loop
lens cleaning cloth for anything other than drying the lens after cleaning.
Approved disinfectant solutions:
• Lysol
Brand III Disinfectant Spray
®
• Lysol
disinfectant or Lysol® concentrate (alcohol-based only)
®
• Cavicide
non-bleach products
• Isopropyl alcohol
• FD 366 (Dürr Dental)
CAUTION: Do not use a metal edged instrument on the OLED display screen.

5.3 CLEANING THE LENS

Inspect the lens after each cleaning. If contaminates are found on the lens or the OLED Display Screen shows a failure
for a Self-Check Dirty Lens Detection, carefully clean with the following method:
1. Clean the lens with a dry Loop
2. Buff the lens surface with your regular disinfectant solution or isopropyl alcohol and a soft cloth using light pressure
in a circular motion. If this does not remove the dental material or contaminant, then proceed to the next step.
3. Use a metal edged (non-diamond tipped) dental instrument to apply lateral pressure to the side and/or edge
of the cured dental material that has bonded to the lens. Take care to not scratch the lens and avoid repeated
scraping motions to clear away the cured dental material.
4. Repeat Steps 1-2. Surface is now ready for use.