Guidance and Manufacturer's Declaration – Electromagnetic Emissions
Loop
is intended for use in the electromagnetic environment specified below.
™
The customer or user should ensure that it is used in such an environment.
Emissions Test
RF Emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
Loop
™
The customer or user should ensure that it is used in such an environment.
Immunity Test
Electrostatic
Discharge (ESD)
IEC 61000-4-2
Electrical fast
+- 2 kV for Power Supply
transient/burst
IEC 61000 4-4
Surge
+- 1 kV differential mode
IEC 61000-4-5
+- 2 kV common mode
<5% U (>95% dip in U) for
Voltage,
dips, shorts,
40% U (60% dip in U) for
interruptions, and
variations on the
70% U (30% dip in U) for
power supply
input lines
>5% U (>95% dip in U) for
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Note – U is the AC main voltage prior to application of the test level.
Compliance
is intended for use in the electromagnetic environment specified below.
IEC 60601 Test Level
+- 8 kV contact
+- 15 kV air
Lines
+- 1 kV for I/O lines
0.5 cycle
5 cycles
25 cycles
5 seconds
30 A/m
Loop
™
function. Therefore, its RF emission are very low
Group 1
and are not likely to cause any interference in
nearby electronic equipment
Class B
Loop
™
including domestic establishments and those
N/A
directly connected to the public low-voltage
power supply network that supplies buildings
N/A
used for domestic purposes.
Compliance Level
+- 8 kV contact
+- 15 kV air
+- 2 kV for Power Supply
Lines
N/A
+- 1 kV differential mode
+- 2 kV common mode
<5% U (>95% dip in U) for
0.5 cycle
40% U (60% dip in U) for
5 cycles
70% U (30% dip in U) for
25 cycles
>5% U (>95% dip in U) for
5 seconds
30 A/m
26
Notes
uses RF energy only for its internal
is suitable for use in all establishments,
Electromagnetic
Environment Guidance
Floors should be concrete
or ceramic tile. If the floors
are covered with synthetic
material, the RH should be at
least 30%.
Mains power quality should be
that of a typical commercial or
hospital environment
Mains power quality should be
that of a typical commercial or
hospital environment
Mains power quality should be
that of a typical commercial or
hospital environment
Power frequency magnetic
fields should be at levels
characteristic of a typical
commercial or hospital
environment