4.2
Operating conditions
Ambient temperature
Relative humidity
4.3
Accessories
The following Cheyenne accessories can
be purchased from your local retailer:
• Cheyenne needle cartridges
• Cheyenne Power Units
• Cheyenne foot switch
• Grips
• Protective hoses
• Cable adapters
• O-rings in two different sizes
Protective hoses must have a diameter of
26 to 31 mm. This corresponds to a hose
width of 40 to 50 mm.
27 EN
5
Getting the HAWK Pen ready
for operation
+10 to +35 °C
30 to 75%
Damage from condensate
If the HAWK Pen is exposed to significant
variation in temperature, e.g. during
transport, condensate can form inside
which can damage the electronics.
Ensure that the HAWK Pen has
reached
before using it. If it was exposed to
high temperature fluctuations, wait
for at least 3 hours for every 10 °C
of temperature difference before
putting it into operation.
Only use the HAWK Pen at ambient
temperatures of +10 to +35 °C.
5.1
Disinfecting equipment
Damage to drive caused by fluid
If disinfectant gets inside the drive, this
can corrode the electrical and mechan-
ical components.
Never immerse the drive in disinfectant.
NOTICE
ambient
temperature
NOTICE
NOTICE
Product damage caused by non-
approved disinfectants
Disinfectants that are not compatible
with the materials of the product can
damage its surface.
Always use disinfectants that are
approved, i.e. that comply with
stipulations in your country.
Note the list of recommended
disinfectants in chapter "Material
compatibilities" on page 32.
Switch off the power supply to the
HAWK Pen.
Check if the HAWK Pen has been
contaminated badly by the return flow
of ink or bodily fluids. In such cases,
perform the steps described in chapter
"Cleaning and maintaining the HAWK
Pen" (see Page 32).
Wipe down the connecting cables, the
drive, the grip and the HAWK Pen Tray
with a soft cloth dipped in disinfectant.
Cheyenne HAWK Pen