commands that have been passed to the interface in the
string. Afterwards the baud rate is also redetermined.
Automatic baud rate detection
The first character that must be sent to the interface
when power on or after the #BD command is a space
(20h). The interface needs it to calculate the baud rate
used for transmission, and automatically adapts it to it. If
any other character or an incomplete character is
received, the system is prevented from working.
DIP switch settings
Nr
On
Off
Function
1
7
8
2
1
2
3
on
off
4
even odd
5
CR
CR+LF
Xon/Xoff protocoll
The command #x1 is used to activate so-called software
handshaking (it is encountered by the command Xo). In
software handshaking mode, data transfer between the
computer and the interface is no longer synchronized by
the hardware lines; instead, this function is performed
by 2 commands:
Xon = 11h = continue transm.
Xoff = 13h = halt transmission
Subject to change without notice
Wordlength
Stop bit(s)
Parity
Parity
End of transmiss.
Command List
(not including commands for the memory card)
Initialization
CLR:
Initializing of the HM8131-2, a stored arbitrary
signal (4k) is retained
RST:
Initializing of the HM8131-2, a stored arbitrary
signal will be replaced by the factory stored one
General Commands
STA?: Queries the instruments configuration
VER:
Queries the software version
SNR?: Queries the serial number
ID?:
Queries the equipment identification (short
version)
*IDN?:
Queries the equipment identification (long version)
TRG:
Trigger activation
STO:x: Store instrument status (0 - 9)
RCL:x: Recalls instrument status (0 - 9)
Bus Commands
LKO:
Recall instrument from "local lockout"
LK1:
Set instrument to "local lockout" mode
RM0:
Re-setting instrument to manual operation
RM1:
Setting instrument to Remote mode
Equipment Adjustments
OT1:
Activates signal output
OT0:
De-activates signal output
RFX:
Activates external reference frequency
RFI:
Activates internal reference frequency
BP0:
Audio tone off
BPS:
Audio tone, medium
BPL:
Audio tone, high
Signal Types
SIN:
Sinusoid
SQR:
Square wave
TRI:
Triangular
RMP:
Ramp, increasing
RMN: Ramp, falling
ARB4: Arbitrary signal, 4 k word length
ARB16: Arbitrary signal, 16 k word length
NOISE: White noise
Signal Parameter
FRQ?:
Queries the adjusted frequency
FRQ:xxx: Set frequency to xxx (preset)
DFR:
Set display to frequency reading
VPP:
Voltage indication in peak to peak
VRMS:
Voltage indication in RMS
AMP?:
Queries actual amplitude
AMP:xxx?: Set amplitude to xxx (preset)
DAM:
Set display to amplitude reading
OFS? :
Queries offset voltage
OFS:xxx : Set offset voltage to xxx (preset)
DOF:
Set display to offset voltage reading
PHA?:
Queries phase value
PHA:xxx: Set phase to xxx (preset)
English