Control panel pushbuttons (keys)

Selection control key for selection of specific parameter.
SELECT
The subsequent parameter is selected by every further keystroke.
Start centrifugation run. The LED in the button lights up during the centrifugation run as long as the
START
rotor is turning.
IMPULS
Short-term centrifugation.
The centrifugation run is effected as long as the button is held down. The LED in the button lights up
during the centrifugation run as long as the rotor is turning.
Store inputs and changes.
End centrifugation run.
STOP
The rotor runs down with a pre-selected brake step. The right-hand LED in the button lights up until the
OPEN
rotor is stationary. Once the rotor is stationary the left-hand LED flashes in the button. Pressing the
button twice triggers the EMERGENCY STOP.
Unlock the lid.
The left-hand LED in the button goes out.
Leave the parameter input.
Switch between RPM and RCF display.
RCF
RCF values are displayed in > <.
Start pre-cooling.
The pre-cooling speed is settable. It is pre-adjusted to 2800 RPM.
13.3

Adjustment possibilities

PROG RCL
Program position of the called-up program.
t/min
Running time. Settable from 0 - 99 min, in 1 min increments.
t/sec
Running time. Settable from 0 - 59 s, in 1 second increments.
Continuous run "∞". Set parameter t/min and t/sec to zero.
RPM
Revolutions per minute. A numerical value from 500 RPM up to the maximum speed of the rotor can be
set. Maximum speed of the rotor, see Chapter "Anhang/Appendix, Rotoren und Zubehör/Rotors and
accessories". Settable in increments of 10.
RAD/mm
Centrifugation radius. Input in mm. For centrifugation radius see Chapter "Anhang/Appendix, Rotoren
und Zubehör/Rotors and accessories". The input of the radius is only possible if the RCF display
(> RCF <) is selected.
RCF
Relative Centrifugal Acceleration. A numerical value can be set, which gives a speed between 500 RPM
and the maximum speed of the rotor. Adjustable up to 100 in intervals of 1, and from 100 in intervals of
10. The RCF value is automatically rounded up or rounded down with regard to the RPM interval. The
input of the RCF is only possible if the RCF display (> RCF <) is selected.
Starting steps 1 - 9. Step 9 = shortest starting time, Step 1 = longest starting time.
Brake steps 0 - 9. Step 9 = shortest run-down time, Step 1 = long run-down time,
Step 0 = longest run-down time (brakeless run-down).
T/°C
Temperature Set Point (only in centrifuges with cooling). Adjustable from -20°C to +40°C, in 1°C
intervals. The lowest obtainable temperature depends on the rotor (see Chapter "Anhang/Appendix,
Rotoren und Zubehör/Rotors and accessories").
PROG STO Program position on which the program is stored. 9 programs can be stored (program positions 1 - 2 - 3 -
... 9). The programme position # serves as temporary storage for altered adjustments.