Trouble Shoo ng/guide for electrical issues with M2 MULTI-
TIP chairs.
1. If only the le /last ba ery light is on.
2. If the 2 smaller lights at the arrow bu ons on
the hand control is on.
HMN a/s
NB! AS A FIRST COURSE OF ACTION
CONCERNING ELECTRICAL ISSUES, AL-
WAYS TRY TO CHARGE THE CHAIR
FOR 24 HOURS!
The chair is low on power and must be charged.
The chair will not elevate if only 1 ba ery light is
on.
The emergency STOP has been ac vated. It must
be deac vated by holding down the emergency
stop bu on on the hand control un l the chair
beeps once and the 2 small diodes at the arrow
keys go out.
3. All 4 lights at the top of the hand control is
blinking.
The chair's control box has received an error sig-
nal. The chair must be reset. This is done by first
pressing the emergency stop bu on on the hand
control un l the emergency stop is ac vated, and
then holding the bu on next to the emergency
stop bu on. First the chair will beep 3-4 mes,
then it will lower the chair completely. When the
chair is sta onary, it is reset.
If the chair a er this is s ll blinking in all 4 top
lights on the hand control it means there is a
defect component.
Contact HMN for further advise.
26
HMN
www.
dk
.