Error message
The evaluation unit is not
reacting to data transfer but
co is shown on the display.
The desired log cannot be set
with the button combination.
The blood pressure monitor
cannot be switched on.
An error occurs during the
first measurement.
Possible cause
1. Data transmission cable has not
been inserted properly into the PC.
2. Also see Err 4
There are still readings from the
last patient in the memory
1. The rechargeable batteries or
batteries have been incorrectly
inserted
2. The power pack or battery
voltage is too low
3. The display is faulty
The cuff size does not fit the
patient's arm circumference.
Troubleshooting
Measures
1. Check whether the 9-pin plug of
the data transmission cable is
securely located in the device's
interface socket.
2. Also see Err 4
Delete the data in the blood
pressure monitor but make sure
that the data has been saved (see
chapter 3.4.2 "Clearing the
memory")
1. Reinsert both rechargeable
batteries or batteries and ensure
correct polarity.
2. Replace the rechargeable
batteries or batteries
3. Send the device to your specialist
supplier or directly to IEM GmbH for
repair.
Check the circumference of the arm
using the tape measure provided
and compare it to the information
printed on the cuff.