Troubleshooting
Error message
Err 9
+
possible continuous alarm
until a button is pressed
Err 10
+
Continuous alarm until a
button is pressed
48
Possible cause
1. Residual pressure inside the cuff
2. Zero point comparison could not
be performed successfully.
1. Serious error due to pressure
build-up outside the measuring
process (pump has been switched
on against orders)
2. These error messages all show
a severe error in the program code.
Measures
1. Wait for the cuff to deflate
completely.
2. Send the device immediately to
your specialist supplier or directly to
IEM GmbH for checking.
Send the device immediately to
your specialist supplier or directly to
IEM GmbH for checking and
repairs.