Care and maintenance
4.2 Maintenance plan
Please check the battery/power pack voltage on a weekly basis.
For the battery/power pack voltage curve, please refer to the HMS CS instructions for use.
Every 2 years:
As proof of continuous compliance to "Basic Requirements" pursuant to Directive 93/42/EEC, the Mobil-O-Graph
blood pressure monitor must undergo metrological checks every two years. In certain countries, this requirement
may be regulated by national laws or regulations.
In addition to re-calibration, IEM products do not require any services referring to low-voltage regulations and
compatibility.
For details on re-calibration services, kindly contact your local IEM Distributor.
Rechargeable batteries:
Rechargeable batteries are liable to ageing. Rechargeable batteries that are damaged, or with which a 24-hour
measurement is no longer possible, must be replaced immediately.
We recommend that you only use the rechargeable batteries supplied by IEM, the capacity and quality of which has
been tested. It is important to make sure that the capacity of the rechargeable batteries is above 1500 mAh.
For further information, please refer to the instructions for use of the charger