Frequency Measuring
range
200.00 µH
2000.0 µH
20.000 mH
10 kHz
200.00 mH
2000.0 mH
20.000 H
Frequency Measuring
range
20.000 µH
200.00 µH
100 kHz
2000.0 µH
20.000 mH
200.00 mH
"L" corresponds to the value read in the unit Henry (H)
No voltages must be applied to the measuring inputs. Discharge all components
before connecting them to the measuring device. Do not touch any circuits or parts
of circuits if they may be subject to voltages higher than 33 V ACrms or 70 V DC!
Danger to life!
60
Resolution Accuracy
0.01 µH
±(0,3%+2)
0.1 µH
±(0,3%+2)
1 µH
±(0,3%+2)
0.01 mH
±(0,3%+2)
0.1 mH
±(0,5%+3)
1 mH
±(2,0%+5)
Resolution Accuracy
0.001 µH
±(0,5%+3)
0.01 µH
±(0,5%+3)
0.1 µH
±(0,5%+3)
1 µH
±(0,5%+3)
0.01 mH
±(1,0%+5)
De
θe
±0,003
±0,17°
±0,003
±0,17°
±0,003
±0,17°
±0,003
±0,17°
±0,005
±0,29°
±0,005
±1,15°
De
θe
±0,005
±0,29°
±0,005
±0,29°
±0,005
±0,29°
±0,005
±0,29°
±0,010
±0,57°
ESR /Rp
±1.88L×10
+2
2
±1.88L×10
+2
2
±1.88L×10
+2
2
±1.88L×10
+2
2
±3.14L×10
+3
2
±1.26L×10
+5
3
ESR /Rp
±3.14L×10
+3
3
±3.14L×10
+3
3
±3.14L×10
+3
3
±3.14L×10
+3
3
±6.28L×10
+5
3