3.4

Set-Up

The Immersion Circulator may be clamped directly to the side of a
reservoir (user supplied) or to a support rod adjacent to the
reservoir. To produce the best fluid circulation, the rotating pump
outlet nozzle should be pointed along one side of the reservoir
wall.
The liquid in the reservoir should be maintained at a depth
sufficient to fully immerse the heater coils, over-temperature
sensor, and pump outlet nozzle.
When using a plastic reservoir, be sure to position the Circulator so
that the heater coils do not contact the sides or bottom of the tank.
3.5

Connections to External Apparatus

Although not designed for closed loop circulation (the Immersion
Circulator does not have inlet and outlet fittings), it may be used for
this purpose.
To connect the Immersion Circulator to an external apparatus, slide
½ inch ID (13mm) tubing over the rotating pump outlet nozzle and
fasten it in place with a hose clamp. Connect the other end of the
tubing to the inlet of the external apparatus. Be sure to run tubing
from the outlet of the external apparatus back to the reservoir to
return the fluid.
3.6

Attainable Temperatures

An Immersion Circulator can be used with reservoirs of various capacities and shapes as well as
with different fluids. These variables may adversely affect temperature accuracy and stability. For
example, a reservoir with large surface area loses heat more quickly, which may prevent the
Circulator from attaining the desired temperature.
The following chart is intended as an approximate guide to temperature performance expectations
under various conditions:
Temperature 30° C
40° C
Water
192L
96L
Oil
283L
202L
Results may vary.
Approximate Attainable Temperatures vs. Liters In Uncovered Reservoir
50° C
60° C
70° C
48L
24L
12L
145L
103L
74L
80° C
90° C
100° C
6L
3L
53L
38L
27L
Minimum
immersion
level
110° C
120° C
130° C
19L
14L
10L
140° C
150° C
7L
5L

Werbung