12.1

Basic data

Accuracy is given as
±([% of reading]+[number of least significant digits])
at 18°C to 28°C, with relative humidity up to 70%.
DC Voltage
Range
0.000...6.599 V
6.60...65.99 V
66.0...659.9 V
660...1000 V
Input impedance: ≥9 MΩ
Overload protection: 1000 V DC or 750 V AC RMS.
AC Voltage (True RMS)
Range
0.000...6.599 V
6.60...65.99 V
66.0...659.9 V
660...750 V
Input impedance: ≥8.5 MΩ
Crest Factor: ≤3.
AC coupled true RMS specified from 5% to 100% of range.
Frequency ranges: 50 Hz ~ 1 kHz. Accuracy of f: ±(0.1% rdg + 5 dgts). The
readings of frequency measurements will show on the sub-display.
Minimum Input voltage Range: >500 dgts.
Overload protection: 1000 V DC or 750 V AC RMS.
12

Technical data

Resolution
0.001 V
0.01 V
0.1 V
1 V
Resolution
0.001 V
0.01 V
0.1 V
1 V
CMP-2000 – USER MANUAL
Accuracy
±(0.5% rdg + 2d)
Accuracy
±(1.5% rdg + 8d)
50...500 Hz
35