7
Technical Data
Sampling principle: Microbial air sampler of
®
the RCS
series employ the impaction principle,
whereby the air stream enters the rotor at
the front of the instrument and the airborne
microbes are separated onto the agar strip
by centrifugal force.
Measurement range: 1-1999 liters (recom-
mended measurement range 10-1000 liters)
Air flow-rate: approx. 50 L/min
Instrument precision: ± 5%
Sample volume: 7 volumes in memory in
ascending order of 10, 20, 50, 100, 200, 500
and 1000 liters. 3 positions in memory, individual
selectable from 1 to 1999 liters. The accuracy
of sampling volume is achieved through cali-
bration.
Rotor speed: approx. 6100 rpm
Maximum allowable axial force on rotor
shaft: 30 Newtons
Power supply: Rapid charging 7.2 V nickel cad-
mium battery (charging time 1 hour)
Automatic power-off: The instrument
switches itself off automatically after
8 minutes
Materials:
• Housing: polycarbonate; resistant to 70%
ethanol
• Rotor: anodized aluminium; autoclavable
• Protection cap: stainless steel; autoclavable
Weight: approx. 1500 g with battery pack
(3.3 Ibs)
Miscellaneous:
• Operation via keyboard panel with integrated
display
• Thread connection for tripod
• Infra-red remote control sensor
• Can be calibrated using the Anemometer
(traceable to PTB, Physikalisch Technische
Bundesanstalt)
• Errors indicated via display and acoustic signal
• Memory function for residual sampling
volume and last sampling volume processed
35