Display Messages/Acoustic Signal

after replacing the battery, and pressing the
[RESET] key two times. If a sampling procedure
has been interrupted, it can now be restarted,
enabling the residual volume shown next to the
colon to be processed.
32
: 075
Sampling is completed;
the instrument has not
yet switched itself off.
000
Sampling is completed;
the instrument has not
yet switched itself off.
LOBAT
Indicates low voltage of
the rechargeable battery.
The message is cleared
Abd
Indicates that the protec-
tion cap is missing or
has not been correctly
installed.
message is cleared by pressing the [RESET] key.
set) indicates that a signal has been received
from the remote control.
Various acoustic signals can sound which indi-
cate either that the sampling process has been
completed or that certain alarm conditions
have been encountered:
Sampling complete
Alarm signal
blc
Indicates that the rotor is
jammed. After rectifying of
the cause of the jam, the
75.0
A decimal point in front
of the last digit (here a
volume of 750 L has been