Installation
4 Installation
4.1 Fixing the rotary vane pump
Tilting danger! Serious injury due to tilting of the product
There is a danger of the vacuum pump not attached tipping over due to changes in the center of
gravity or incorrect loading. Serious injuries due to trapping or crushing of limbs, e.g. feet, are the re-
sult.
► Do not use the vacuum pump as a climbing aid.
► Do not exert any force on the product.
► Ensure that the product has a safe centre of gravity, when mounting components.
► Wear protective equipment , e.g. protective shoes
Fig. 4:
1 Base frame
Fixing the rotary vane pump
► Screw the rotary vane pump onto the anti-vibration buffers on a fixed, horizontal base.
4.2 Installing vacuum pump combinations
Rotary vane pump
Hena 202 K
Hena 302 K
Hena 401 K
Hena 631 K
Tbl. 1:
24/34
WARNING
2
Fixing the rotary vane pump
2
Anti-vibration buffer (4x)
Installation conditions
Observe the installation conditions for the roots pump and the rotary vane pump. For de-
tails, see the respective operating instructions of the vacuum pump. The weight of the roots
pump must not exceed the value specified in the following table.
Exhaust
G2 female
G2 female
G3 female
G3 female
Installation conditions for vacuum pump combinations
Intake flange
max. permissible weight of the roots pump
DN 100 ISO-F
300 kg
DN 100 ISO-F
300 kg
DN 160 ISO-F
600 kg
DN 160 ISO-F
600 kg