Close the lid of the centrifuge. Now proceed as described previously to set all important run parameters. If the lid
isn't closed when storing the program, in the display "RPM/RCF" (A-1), the words "FirSt" and "CLOSE Lid" (see
figure 20) flashes alternately. If you want to start the run without storing the program, in the display "RPM/RCF" (A-
1), the words "First" and "PrESS StoreE" (See figure 21) flashes alternately.
For adaption of data press the key "Store" (2) (refer to figures 20 and 21) for approx. 1 second. If the program is
stored correctly, the word "StorE" appears in the display "RPM/RCF" (A-1). As a result, the word "programm" (M10)
disappears.
As soon as the key "Store" (2) is released, the word "program xx" (M10) reappears – the (xx) stands for the chosen
program location.
If all program numbers are occupied, you can take an old number that is not necessary anymore and just put in the
new parameters.
3.5.2
Recall of stored programs
To recall stored programs, press the key "Prog/Setup" (3) (refer to figure 22) while the lid is already closed. Inside
the display "Time/Prog" (A-4), "programm –"(M10) appears. The desired program number can be pre-selected
with the adjusting dial (4).
In the respective displays the stored values for that program will appear.
If the wrong rotor is inside the centrifuge for the pre-selected program, in the display "RPM/RCF" (A-1), the word
"rotor" (M3) flashes. At the same time the word "FALSE" and the stored rotor number " xx" (M4) will flashing by
turns.