Use With The Half Duplex Adapter; Remote Command Names; List Of Remote Commands - sylvac UL III Bedienungsanleitung

Extraleichte messschieber
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

The data output can be requested from the computer by toggling the DTR line in the following
modes:
See also user manual of OPTO-RS.
VIII.5 Use with the half duplex adapter
With the half duplex adapter, you may completely control the instrument from the computer by
sending remote commands in the following modes
See OPTO-RS user manual for connexion on a computer
VIII.5.1 Remote command names
Most commands consist of 3 characters followed by either a 0 or 1 (disabled or enabled). Note:
commands may be upper or lower case and should be followed by the ASCII code <CR>.
The instrument echoes the command.
VIII.5.2 List of remote commands
<NOR>
Will set the instrument in the measure mode (or in reference mode if the button is disabled)
<MOD?>
The instrument returns its current mode of operation (NOR, REF, MIN,MAX, DEL, TOL1)
<STO0>, <STO1>
Deactivates or activates the measurement "HOLD"
<KEY0>, <KEY1>
Deactivates or activates [mode] button
<RST>
Reset the instrument to the factory default condition
<SET?>
The instrument returns its current parameters (MM RES2 REF1 etc)
Note:
- B1 Battery OK
- B0 Battery must be replaced
<ID?>
The instrument returns its class identification
- SYxxx: xxx product number
- measuring
- references
- minmax
- tolerances
- measuring
- references
- minmax
- tolerances