Centrifuga FC5706
12.2 Tabella 2: Peso netto ammesso
12.3 Tabella 3: Velocità max. e valori RCF per rotori ammissibili
12.4 Tabella 4: Tempi di accelerazione e decelerazione
12.5 Tabella 5: Messaggi di errore
Error-no.:
1
2
14
30
33
55
60
70
Rotor-number
30130880 6 x 5 ml swing
30130877 12 x 15 ml
30130878 6 x 50 ml
Rotor-number
30130880 6 x 5 ml swing
30130877 12 x 15 ml
30130878 6 x 50 ml
Rotor-number
30130880 6 x 5 ml
swing
30130877 12 x 15 ml
30130878 6 x 50 ml
Description
Imbalance
Imbalance sensor is defective
Leap of speed is to big between two measurments
Motor is blocked or detective
Open lid during the rotor is running
Overspeed
Undervoltage in the intermediate circuit
sticking relay
Max. speed
-1
4000 min
-1
6000 min
-1
6000 min
Max. speed
-1
4000 min
-1
6000 min
-1
6000 min
Acceleration values
Deceleration values
level 0
level 9
35
8
95
33
91
41
in seconds
acceleration time
-1
from 0 min
-> V
from V
max
Permissible
net weight
240 g
300 g
432 g
value
1878 x g
4427 x g
4427 x g
level 0
level 9
25
7
257
51
274
52
deceleration time
-1
-> 0 min
max
IT-21