Herunterladen Diese Seite drucken

OHAUS Frontier FC5706 Bedienungsanleitung Seite 13

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 81
TM
Frontier
Centrifuge FC5706
3.3
Lid control
3.3.1
Lid open
After the run, when closing the lid of the centrifuge, in the display "RPM | RCF" (A-1) the word "close" (M1)
appears (refer to figure 18 below).
If there is a rotor in the centrifuge, additional the word "rotor"(M3) appears, as well as the code number of the
respective rotor which is in the centrifuge system "77" (M4). If there is no rotor in the centrifuge, the word "rotor"
(M3) flashes and additional the word "no" (M4) appears. By pressing the key "Door Open" (7) you can release the
lid of the centrifuge. As soon as the electromagnetic lid is completely released, the word "open" (M2) appears.
Now you can open the lid of the centrifuge.
A-1
During the run you can call up the rotor type at any time by pressing the key "Door Open" (7).
3.3.2
Lid lock
The lid should only be put down slightly. An electromagnetic lid lock closes the lid, at the same time the word
"open" (M2) disappears (refer to figure 18).
As a sign that the centrifuge is ready for starting, in the display "RPM | RCF" (A-1) the word "close" (M1) appears.
Simultaneously the word "rotor" (M3) is displayed, as well as the code number of the rotor, which is in the
centrifuge system, "no 71" (M4). With that, all rotor specific data, like e. g. max. speed, acceleration etc., are
adopted.
ATTENTION:
Before closing the lid please check if the rotor is tighten, and all 6 buckets have put in the swing out
rotor.
M2
M3
M1
11
Figure. 18
EN-7
7

Werbung

loading