Recommended separation distances between portable and mobile RF communications equipment and the
EN
EQUIPMENT or SYSTEM - for EQUIPMENT and SYSTEMS
The Theragun PRO PLUS is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Theragun PRO
PLUS can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Ther-
agun PRO PLUS as recommended below, according to the maximum output power of the communications equipment
Rated maximum output of transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
8
Recommended separation distances between
portable and mobile RF communications equipment and the Theragun PRO PLUS
150 kHz to 80 MHz
outside ISM and amateur radio bands
0.12
0.38
1.2
3.8
12
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
in ISM and amateur radio bands
0.20
0.63
2.00
6.32
20.00
80 MHz to 800 MHz
800 MHz to 2.7 GHz
0.035
0.11
0.35
1.10
35
0.07
0.22
0.70
2.21
70