Guidance and manufacturer´s declaration – electromagnetic immunity –
for EQUIPMENT and SYSTEM
The Theragun PRO PLUS is intended for use in the electromagnetic environment specified below. The customer or the user of the Theragun PRO PLUS should assure that it is used
Immunity test
IEC 60601 test level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
6 V in ISM and amateur radio bands
between 0,15 MHz and 80 MHz
Radiated RF
10 V/m
IEC 61000-4-3
80 MHz to 2.7 GHz
385MHz-5785MHz Test specifications
for ENCLOSURE PORT IMMUNI-
TY to RF wireless communication
equipment (Refer to table 9 of IEC
60601-1-2:2014)
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic is affected by absorption and reflection from structures, objects and people.
a The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6,765 MHz to 6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz. The amateur radio bands
between 0,15 MHz and 80 MHz are 1,8 MHz to 2,0 MHz, 3,5 MHz to 4,0 MHz, 5,3 MHz to 5,4 MHz, 7 MHz to 7,3 MHz, 10,1 MHz to 10,15 MHz, 14 MHz to 14,2 MHz, 18,07 MHz to 18,17 MHz, 21,0 MHz to 21,4 MHz, 24,89 MHz to 24,99
MHz, 28,0 MHz to 29,7 MHz and 50,0 MHz to 54,0 MHz.
b
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which theTheragun PRO PLUS is used exceeds
the applicable RF compliance level above, the Theragun PRO PLUS should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re¬orienting or relocating
theTheragun PRO PLUS.
c Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.
Guidance and manufacturer´s declaration – electromagnetic immunity
in such an environment.
Compliance level
Portable and mobile RF communications equipment should be used no closer to any part of the Theragun PRO PLUS,
including cables, than the recommended separation distance calculated from the equation applicable to the frequency of
the transmitter.
Recommended separation distance
N/A
10 V/m
80 MHz to 2.7 GHz
385MHz-5785MHz Test
specifications for ENCLOSURE
PORT IMMUNITY to RF wireless
communication equipment
(Refer to table 9 of IEC 60601-
where p is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer
1-2:2014)
and d is the recommended separation distance in metres (m).b
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,a should be less than the
compliance level in each frequency range.b
Interference may occur in the vicinity of equipment marked with the following symbol:
Electromagnetic environment -guidance
EN
7