Guidance and manufacturer's declaration – electromagnetic immunity –
EN
for all EQUIPMENT and SYSTEMS
The Theragun PRO PLUS is intended for use in the electromagnetic environment specified below. The customer or the user of the Theragun PRO PLUS should assure that it is used
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrostatic
transient / burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short interruptions and voltage
variations on power supply input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
6
Guidance and manufacturer´s declaration – electromagnetic immunity
IEC 60601 test level
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV,
± 15 kV air
± 2 kV for power
supply lines
± 1 kV for input/output lines
± 1 kV differential mode
± 2 kV common mode
0 % UT; 0,5 cycle g) At 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°
0 % UT; 1 cycle and 70 % UT; 25/30 cycles Single phase: at 0°
0 % UT; 250/300 cycle
30 A/m
NOTE
UT is the a. c. mains voltage prior to application of the test level.
in such an environment.
Compliance
level
± 8 kV contact
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic
± 2 kV, ± 4 kV, ± 8 kV,
material, the relative humidity should be at least 30 %.
± 15 kV air
N/A
Mains power quality should be that of a typical commercial or home environment.
N/A
Mains power quality should be that of a typical commercial or home environment.
Mains power quality should be that of a typical commercial or home environment. If the user of
the Theragun PRO PLUS requires continued operation during power mains interruptions, it is
N/A
recommended that the Theragun PRO PLUS be powered from an uninterruptible power supply
or a battery.
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical
30 A/m
commercial or home environment.
Electromagnetic environment -guidance