Hach 3400 sc Bedienungsanleitung Seite 71

Digitales leitfähigkeits- analyse-system

Werbung

Group Name
Setup/Relay 2/Timer
Setup/Relay 2/Timer
Setup/Relay 2/Timer
Setup/Relay 2/Status
Setup/Relay 3
Setup/Relay 3
Setup/Relay 3
Setup/Relay 3
Setup/Relay 3
Setup/Relay 3/Alarm
Setup/Relay 3/Alarm
Setup/Relay 3/Alarm
Setup/Relay 3/Alarm
Setup/Relay 3/Alarm
Setup/Relay 3/Alarm
Setup/Relay 3/Control
Setup/Relay 3/Control
Setup/Relay 3/Control
Setup/Relay 3/Control
Setup/Relay 3/Control
Setup/Relay 3/Control
Setup/Relay 3/Control
Setup/Relay 3/Event
Setup/Relay 3/Event
Setup/Relay 3/Event
Setup/Relay 3/Event
Tabelle 16 Controller ModBUS Registers (Fortsetzung)
Tag Name
Register # Data Type
Duration Time
40126
Period Time
40127
Off Delay
40128
Level
40129
Source
40130
Sensor Select
40131
Measurement
40132
Select
Type
40133
Transfer
40134
Setting
High Alarm
40135
Low Alarm
40137
High
40139
Deadband
Low
40141
Deadband
On Delay
40143
Off Delay
40144
Setpoint
40145
Phase
40147
Deadband
40148
Overfeed
40150
Timer
On Delay
40151
Off Delay
40152
Reset
Overfeed
40153
Timer
Setpoint
40154
Phase
40156
Deadband
40157
On Max Time
40159
Length
R/W
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Float
2
R/W
Float
2
R/W
Float
2
R/W
Float
2
R/W
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Float
2
R/W
Unsigned
1
R/W
Integer
Float
2
R/W
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Float
2
R/W
Unsigned
1
R/W
Integer
Float
2
R/W
Unsigned
1
R/W
Integer
ModBUS Register Information
Description
Sets the timer on duration time (seconds)
Sets the period between timer on events
(minutes)
Sets the time the affected sensor outputs
are held/transferred after the timer turns
off (seconds)
Sets the status level which will trigger the
relay
Selects data source for this relay
(0 = None; 1 = Real Time Clock;
2 = Sensor; 4 = Calculation)
Selects sensor source when source =
Sensor (0 = Sensor1; 1 = Sensor2)
Selects measurement on the sensor
(0 = Meas1 .. 3 = Meas4)
Selects the relay type (0 = Alarm; 1 =
Control; 2 = Status; 3 = Timer; 4 = Event)
Selects the transfer value for the relays
(0 = De-energized; 1 = Energized)
Sets the high alarm setpoint
Sets the low alarm setpoint
Sets the high alarm deadband
Sets the low alarm deadband
Sets the on delay time
Sets the off delay time
Sets the controller setpoint
Sets the controller phase
(0 = Low; 1 = High)
Sets the controller deadband
Sets the overfeed timer value (minutes)
Sets the on delay time (seconds)
Sets the off delay time (seconds)
Resets the overfeed timer
Sets the event setpoint
Sets the event phase (0 = Low; 1 = High)
Sets the event deadband
Sets the max on time (minutes)
69