Herunterladen Diese Seite drucken

Hach 2100N IS Grundlegende Bedienungsanleitung Seite 11

Vorschau ausblenden Andere Handbücher für 2100N IS:

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 17
7. Remove the vial
8. Do steps 5–10 for
from the sample cell
the other StablCal vials
holder.
(from lowest to highest
NTU standard).
The S0 light turns on
after the last vial is
measured.
StablCal standards storage
• Do not move a StablCal standard to a different container for storage.
Keep StablCal standards in the plastic case provided with the cover
closed.
• Store at 5 to 25 °C (41 to 77 °F).
• For long-term storage (more than one month between use), keep at
5 °C (41 °F).
Turbidity measurement
W A R N I N G
Potential explosion and fire hazard. This instrument is for measuring water based
samples. Do not measure solvent or combustible based samples.
For accurate turbidity readings use clean sample cells and remove air
bubbles.
Measurement notes
Proper measurement techniques are important in minimizing the effects
of instrument variation, stray light and air bubbles. For accurate and
repeatable measurements:
9. Push CAL.
The instrument saves
the new calibration data
and goes back to
Measurement mode.
Instrument
• Make sure that the instrument is on a level, stationary surface that is
free of vibration during the measurement.
• Instrument stabilization is immediate. No warm-up time is necessary.
• Always close the sample compartment lid during measurement,
calibration and storage.
• Remove the sample cell from the instrument and turn off the
instrument if the instrument is stored for an extended time period
(more than a month).
• Keep the sample compartment lid closed to keep dust and dirt out.
Sample cells
• Always cap the sample cell to prevent spillage of the sample into the
instrument.
• Always use clean sample cells in good condition. Dirty, scratched or
damaged cells can result in readings that are not accurate.
• Make sure that cold samples do not "fog" the sample cell.
• Store sample cells filled with distilled or deionized water and cap
tightly.
English 11

Werbung

loading