Basic error codes:
AS 130.1 Universal attachment
g
2000
1750
1500
1250
1000
750
500
250
0
0
80
160
240
320
400
g
2500
2000
1500
1000
500
0
0
80
160
240
320
400

Error codes

ATTENTION!
If an error code is encountered, an audio warning
signal is generated and the red Off signal light goes on in the display
field. First, please try to continue operation by turning the device off
and back on again.If it is still not possible to eliminate an error after
an extended wait, please contact our service department. If you do so,
you should always tell us what error code has been encountered. This
simplifies the process of tracking down the error and makes it possible
to form a preliminary opinion.
1/min
480
560
640
720
800
1/min
480
560
640
720
800
Error code
Error
Off - LED (red)
The potentiometer for
the speed is not assig-
1x flash /
ning any target value
2 sec pause
Off - LED (red)
Permissible motor tem-
peratur exceeded
2x flash /
2 sec pause
Motor locked
Read fork signal not
available
Off - LED (red)
Maximum adjustable
speed of the device
3x flash /
exceeded
2 sec pause
Off - LED (red)
Motor standstill for de-
vice switch-on proce-
4x flash /
(safety confirma-
2 sec pause
dure
tion not ensured
Off - LED (red)
Safety relay cannot
be activated
5x flash /
2 sec pause
Off - LED (red)
The switch for the TIME
function is not assigning
6x flash /
any target value
2 sec pause
Off - LED (red)
Problem with the power
supply
7x flash /
2 sec pause
Off - LED (red)
Triac cannot be activated - Internal device error
8x flash /
2 sec pause
Cause of the error
- Too rapid a change in the
speed target value
- Internal device error
- Permissible supported
weight exceeded
- Permissible ambient
temperature exceeded
- Agitation table is being
hindered in this stroke mo-
tion by an outside effect
- Internal device error
- Resonance behavior of
the experiment setup
- When turning on the
device, you must ensure
that the motor is not
performing any rotational
motion
- Internal device error
- Internal device error
- The device is beeing
operating with under vol-
tage or overvoltage (for
permissible voltage range,
see technical data)