Control error codes:
ATTENTION!
If an error code is encountered, an audio warning
signal is generated and the red Off signal light goes on in the display
field. First, please try to continue operating by turning the device off
and back on again. If it is still not possible to eliminate an errorafter
an extended wait, please contact our service department. If you do so,
you should always tell us what error code has been encountered. This
simplifies the process of tracking down the error and makes it possible
to form a preliminary opinion.
Error code
Error
Er 2
No communication bet-
ween PC and control de- - PC is not sending any
vice in remote operation
with active watchdog func-
tion in Mode 1
Er 3
Internal device tempe-
rature too high
Er 4
Motor locked or over-
loaded, Problem with
read fork signal
Er 9
Error while reading
stored values.
Er 41
Triac defective
Er 42
Safety relay defective
WD
No communication bet-
ween PC and contol de-
vice in remote operation
with active watchdog func-
tion in Mode 1
30

Interface and outputs

(KS 130 control version only)
The KS 130 control is equipped with a 15-pin SUB-D connector on the rear side
of the device. The pins are assigned in operatin with analog and serial signals.
Analog autput
Voltage values for the speed measurement quantify
are present on the analog pins.
(10)
(15)
Cause of the error
RS 232 C serial interface
- Interface not connected
The serial assignment of the socket can be used to control the
device externally by means of a PC and a suitable application
data within the amount
program.
of time set by the watch-
Configuration of the serial RS 232 C interface
dog
• The function of the interface line between the laboratory device and the auto-
- Permissible ambient
mation system is a selection of the signals specified in EIA
temperature exceeded
Standard RS 232 C, corresponding to DIN 66020 Part 1.
- Agitator table is being
For the assignment of the signals, please refer to the illustration.
hindered in its stroke motion
• Standard RS 232 C applies to the elctronic properties of the
- Internal device error
interfaces and the assignment of signal states in accordance
- BLP Logik
with DIN 66259 Part1.
• Transmission procedure:
- Internal device error
- Internal device error
• Type of transmission:
- Interface not connected
• Character format:
- PC is not sending any
data within the amount
of time set by the watch-
dog
• Transmission speed:
• Data flow conrol:
• RTS:
• RTS:
Analog GND
Speed measurement value 1VDC / 1000 rpm
Asynchronous character
transmission in start-stop
mode
full duplex
Character creation according
to the data format in DIN 66022 for start-stop
mode. 1start bit; 7character bits; 1parity bit
(even); 1stopbit.
9600 baud
RTS/CTS hardware handshake
(Pin 7) LOW (positive tension) / PC may send
(Pin 7) HIGH (negative tension)
/ PC may not send
Analog GND
1V = 1000 1/min
RTS
RxD
TxD
CTS
Analog GND
1V = 1000 1/min
RS232_GND