The device is maintenance-free. It is only subject to the natural
wear and tear of components and their statistical failure rate.
Cleaning:
Disconnect main plug prior to cleaning!
Use only cleaning agents which have been approved by
IKA
®
to clean IKA
®
devices.
Dirt
Cleaning agent
Dye
Isopropyl alcohol
Construction
Water containing tenside/isopropyl alcohol
material
Cosmetics
Water containing tenside/isopropyl alcohol
Foodstuff
Water containing tenside
Fuel
Water containing tenside
For materials which are not listed, please request informa-
tion from IKA
application support.
®
Wear protective gloves while cleaning the device.
Electrical devices may not be placed in the cleansing agent for
the purpose of cleaning.
Do not allow moisture to get into the device when cleaning.
If a different cleaning or decontamination method than the
method defined by IKA
®
with IKA
that this method does not damage the device.
®
Any malfunctions during operation will be identified by an error message on the display.
Proceed as follows in such cases:
 Switch off device using the main switch at the back of the device.
 Carry out corrective measures.
 Restart device.
Error code Effect
Err 10
Stirring function doesn't start.
Err 14
Stop stirring or change according to
relating setting.
If the actions described fails to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the service department
- Send the device for repair, including a short description of the fault.
• IKAFLON
Magnetic stirring bar
®
• TRIKA
®
Magnetic stirring bar
• RSE Stirring bar remover
See more accessories on www.ika.com.

Maintenance and cleaning

is planned, the user must ascertain
Spare parts order:
When ordering spare parts, please give:
- Machine type.
- Serial number, see type plate.
- Item and designation of the spare part,
see www.ika.com, spare parts diagram and list.
Repair:
Please send the device for repair only after it has
been cleaned and is free from any materials which
may constitute a health hazard.
For repair, please request the "Decontamination Certifi-
cate" from IKA
website www.ika.com.
Return the device in its original packaging. Storage packag-
ing is not sufficient. Also, please use suitable shipping pack-
age materials.

Error codes

Cause
Speed setting signal error.
PC communication failure

Accessories

• PC 1.1 Cable
• Labworldsoft
®
, or download printout of it from the IKA
Solution
- Contact IKA
service department.
®
- Check communication cable.
®
®