Operating modes

Settingthe operation mode (control model only)
The Control device model can be operated in the modes described
below:
Operation mode A:
The KS 130 control is switched into operating mode A when it leaves
the factory. After the device is turned on with the power switch (flip
switch), the agitation and time expiration functions are turned off.
The set target values are stored and are displayed digitally. They are
adopted when the corresponding functions are turned on. the set
target values can be varied. After the power has been turned off and
back on, both functions must be started again. The upper speed limit
can only be set or changed in operating mode A. You can recognise
operating mode A by the factual no star is displayed in the menu bar
display of the time setting.
After the power has been interrupted, the device will no longer auto-
matically start up in operating mode A by itself any more.
After you have selected the operating mode and turned on the device,
the following process will be shown on the display.
a.) All display elements in the digital displays are activated for about 2
seconds.
b.) The device type (left display) and operating mode (right display) are
displayed for about 2 seconds.
I 3 0
K S
c.) After this, the set (ore stored) upper speed limit of the agitator
is shown in the left display with SAFE, the unit of measure rpm and
flashing SV for about 2 seconds. As long as the display is flashing
with SV , it is possible to change the upper speed limit by pressing
the turn button and turning at the same time.
2 3 0
S A F E
I
/
m i n
SV
d.) If the upper speed limit is fixed, the unit 1/min in the left display
changes to OK. The new time target value that is set or stored in the
right display (for example 3:58) is displayed with an indication of h
min and a flashing SV .
2 3 0
S A F E
O K
SV
e.) The new speed target value that has been set or stored in the left
display (for example 130) is then shown with the unit 1/min and
flashing SV .
I 3 0
I
/
m i n
SV
f.) By pressing the left or the right turn button, you can place the agi-
tator in operation with the stored value or the one that has just been
set. The actual speed and time values are then shown in large format
on the display. The actual value of the time setting must be understood
A
from this point onward as the time remaining for the agitation process.
If zhe remaining has reached zero, the agitator stops its motion.
A
3 5 8 :
i
n
m i n
SV
3 5 8 :
i
n
m i n
SV